Hardware Problems, Symptoms & Fixes

Symptom 1: Coil not recognized by MR Scanner

Cause: Bad solder joint for wire going to coil ID board

Fix: Re-soldering the joint

Symptom 2: Rattling in the Gateway

Cause: Disconnected 50 Ohm terminator in the gateway

Fix: Resoldering the disconnected terminator

Cause: LNA becoming disconnected

Fix: Reconnecting LNA, optional zip tie fix

Alternative Cause: Dead LNA

Alternative Fix: Swapping out LNA altogether

Symptom 3: No Signal

Symptom 4: Poor signaling/shading

Cause: Broken/sheared BNC cable in the coil

Fix: Replace BNC cable

Symptom 5: Intermittent Loss of Signal

Cause: Dislodged decoupling circuit

Fix: Retuning, soldering, and curing trap

Symptom 6: Low/no signal at start of acquisition with short TEs and moderate flip angles

Cause: Slow response of LNA (preamp)

Fix: Modified/faster LNA from MR Clinical Solutions

Alternative Fix: Increase tune delay for coil switching